    Specification

    Electrical Requirements

     a.c.198~242V, 50/60Hz, 200VA

    Operation Interface

     Selectable in English, French, German and Chinese

    Flow Rate

     0.1 mL/sec. to 10 mL/sec. in 0.1 mL/sec.

    increments Syringe Capacity

     115ml for Shine-C11; 200ml for Shine-C22

    Volume

     0.1 ml to syringe capacity in 0.1 ml increments

    Characteristic

    ● Shine-E Dual Syringes CT Injector provides saline flush capability to your CT. Saline flush not only can control the contrast agent to reach the right part punctually but also may reduce the picture interferential shadows during the breast scan. And the press makes the contrast agent denser so that you can obtain a stronger signal. 

    ● Large color touch-screen makes operation more intuitive and makes injection planning and process controlling easier.

    ● Clear and intelligible operating screen.

    ● Flexible injection design can help you realize the simplest 1 phase of single syringe injection, or finish the complicated 8 phases of dual syringe injection. All are in your control. 

    ● The elaborate assistant tray on which the operator can place the fluid bottles or other things, and put the connector tubing or the scalp vein set in the special place.

    User-Friendly  Operations

    ● 12” wide touch screen console, capable of setting 8 different  injecting phases easily 

    ● All operations can be performed on the console and through the injector arm control

    ● The LED displayer on  the injector head can display in real-time the amount of contrast medium 

    ● Smart operation keys, designed to complete all functions by minimal keying

    Reliability & Safety

    ● Accurate control injecting speed and volume

    ● Comprehensive real-time monitoring of the pressure and system operating states

    ● Remote control of injection to reduce the radiation harmful to the operators

    ● Air-extraction warning before injections, reduce misoperation

    Scientific & Humanized Design

    ● The injector arm's height and direction are adjustable.

    ● The pedestal can be moved and locked randomly at the expected place.

    ● The accessary tray frees the user's hands to focus on the operation.

    Real-time injection pressure monitoring

    With the highly sensitive pressure transducers, it can display the pressure waveforms of injections, control the pressure in the pre-set ranges, monitor the pressure derivatives, control the injecting procedure, and effectively reduce the leakage and save the contrast medium.

    Saline Flush and Dilution

    Saline flush can form a much more denser contrast media bolus. It will boost the accuracy of contrast injecting time, reduce the waste of contrast substance, and diminish the imiging artfacts in the tcrso scan.

    In coronary imaging, diluting saline injection can make the ventricles and coronary arteries have consistent contrast densities. Which can boost the image clarity with minimal artifacts.

    Dual Flow with Setting Motion Ratios

    Shine-E dual CT injector system option is more than a saline flush after a contrast boll the operator can inject both contrast and saline at the same time by setting the motion rat With the proper ratio, left and right heart ventricles can be illuminated uniformly or imroved image quality minimal artifacts.

    Flexible 8 phases injection setting protocols

    Clear and simple console operation, flexible &custamized protocol setting interface.

    Users can easily set one-phase injection or complex 8-phase injection as well to meet the needs of all clinic applications.

    Specification

    Technical specification

    Description

    Electrical Requirements

    220VAC/110VAC,50 or 60Hz,200VA

    Operation Interface

    Available in English, French, Spanish, German and Chinese

    Flow Rate

    0.1-10 m/sec.in increments of 0.1 ml/sec.

    Syringe Capacity

    115ml for Shine-C11;200ml for Shine-C22

    Volume

    0.1 ml to syringe capacity in 0.1 ml increments

    Pressure

    Preset limit:300 psi default, user settable from 50 to 350 psi in 1 psi increments

    Display: Real-time pressure curve

    Change threshold: Stop injection if pressure changes suddenly

    Alarm:Voice alarm and stop injection when over preset limit

    Injection or Scan Delay

    to 3599 seconds in 1 sec, increments

    Autofil

    Settable flow rate from 3-8ml/sec.,volume from 10ml to syringe capability

    Injection Capabilties

    Programmable up to 8 phases

    Pause Phase

    0~999 sec. in 1 sec.increments

    Hold Capability

    30 minutes max

    Storage Capabilities

    1000 protocols

    Expel air locked

    Unable to inject unless expelling air

    Test injection

    Settable flow rate from 0.1 -5ml/sec.,volume from 0.1ml to 10ml

    Heat maintainer

    35-38℃(optional)

    Hand switch

    Hand switch control pause and continue injection (optional)

    Emergency stop function

    When the system is injecting, click on any point of the touch screen or any button on the injector arm,the system will stop the injection

    Dual Flow

    Settable in a ratio of saline and contrast

    KVO

    Independent KVO program

    Dimensio

    Injector Arm and Pedestal:L:22.8(58.0cm),W:12.5(20.1cm),H:50.2“(127.4cm)

    Console:L:4.55(11.55cm).W:12.68(31.8cm),H:9.77“(24.81cm)

    Weight

    Injector Arn and Pedestal:19.5Kg

    Console:4.1Kg

    Suspension system of injector arm (Optional)

    ● Save space in the scanning room

    ● Operation more flexible

    ● Injector arm easier move to the right position

    P/N C01-006-10